IM 701510-01E
16-8
16.10 General Specifications
Item
Specifications
Safety standard*
Complying standard
:EN61010
• Overvoltage category II
• Pollution degree 2
Emission*
Complying Standard
: EN55011:Class A, Group1
This is a Class A product(for industrial environment). In a domestic environment, this product may
cause radio interference in which case the user may be required to take adequate measures.
• Cable conditions for DL1540/DL1540L
• External trigger/clock input terminal
Connected BNC cable(3D2W), with both side’s Ferrite Cores(TDK:ZCAT2035-0930A)
• Trigger output terminal
Same as above for the External trigger/clock input terminal
• Connecter for half pitch interface
Connected the cable B9920TA(the dedicated cable for RS-232-C Unit-700927)
• Cable conditions for DL1520/DL1520L
• Connector for RS-232-C interface (for suffix code -C2)
Connected RS-232-C shielded cable, with both side’s Ferrite Cores(TDK:ZCAT2035-0930A)
• Connector for Centronics interface (for suffix code -C3)
Connected the cable B9916TB(the dedicated cable for suffix code -C3)
• Connecter for half pitch interface (for the SCSI interface, on DL1520L with suffix code -C4)
Use a cable less than 3 m long with a characteristic impedance between 90 and 132
Ω
. Also,
attach a ferrite core (TDK:ZCAT2035-0930A) on each end of the cable.
Immunity*
Complying Standard
: EN50082-2:1995
• Susceptibility under immunity condition
Noise increase
: <
±
15 mV
SCSI interface
: No effect
• Testing Condition
100 Msps 5 mV/div PeakDetect(Envelope)Mode 20 MHzBWL, Channel Input terminated with
50
Ω
.
Standard operating conditions
Ambient temperature
:
23
±
2
°
C
Ambient humidity
:
55
±
10 % RH
Power voltage and frequency fluctuation : Less than 1 % of the rated voltage/frequency
Warm-up time
30 min. or more
Storage conditions
Temperature
: –20 to 60
°
C
Humidity
: 20 to 85 % RH (no condensation allowed)
Operating conditions
Temperature
: 5 to 40
°
C
Humidity
: 20 to 85 % RH (without printer)
: 35 to 85 % RH (with printer) (No condensation allowed in either case)
Storage altitude
3000 m or below
Operating altitude
2000 m or below
Rated supply voltage
DL1540, DL1540L (for suffix code -1 and -5) : 100 to 120 VAC/200 to 240 VAC
DL1520, DL1520L (for suffix code -1)
: 100 to 120 VAC
DL1520, DL1520L (for suffix code -5)
: 200 to 240 VAC
Permitted supply voltage range
DL1540, DL1540L (for suffix code -1 and -5) : 90 to 132 VAC/180 to 264 VAC
DL1520, DL1520L (for suffix code -1)
: 90 to 132 VAC
DL1520, DL1520L (for suffix code -5)
: 180 to 264 VAC
Rated supply voltage frequency
50/60 Hz
Permitted supply voltage frequency
48 to 63 Hz
Power Supply Fuse
DL1540, DL1540L (for suffix code -1 and -5) : 250 V, 3.15 A, Time lag, VDE/SEMKO approved
DL1520, DL1520L (for suffix code -1)
: 250 V, 3 A, Time lag, UL/CSA approved
DL1520, DL1520L (for suffix code -5)
: 250 V, 2 A, Time lag, VDE/SEMKO approved
Maximum power consumption
DL1540, DL1540L (for suffix code -1 and -5) : 220 VA/280 VA
(when the built-in printer is used)
DL1520, DL1520L (for suffix code -1)
: 130 VA
DL1520, DL1520L (for suffix code -5)
: 140 VA
(when the built-in printer is not used; refer to page 3-5)
* Applies to products manufactured after Jan. 1997 having the CE Mark. For all other products, please contact your nearest YOKOGAWA