9-1
IM 701510-01E
Performing a GO/NO-GO
Action
9
9.1
Judging using a Waveform Zone
Operating Procedure
1.
Press the
GO/NO-GO key
to display the GO/NO-GO setting menu.
2.
Press the “
Mode
” soft key and select “
ON
”. The GO/NO-GO top menu appears.
3.
Press the “
ZONE
” soft key. The default setting is “
ZONE
”.
When using the GO/NO-GO determination funtion, the following settings/selections can be
specified.
• Creating the determination zone (refer to page 9-1, 9-2);
• Saving the determination zone (refer to page 9-3);
• Redrawing/editing the saved determination zone (refer to page 9-3);
• Setting up the determination conditions (refer to page 9-4, 9-5);
trace conditions, determination logic, determination sequence, action after determination
• Executing and aborting the determination function (refer to page 9-5).
Note
The determination waveforms are saved in the internal memory. Old determination waveforms which are
already present in the internal memory, will be overwritten when new ones will be saved. An error
message will not appear. Therefore, verify before saving whether or not determination waveforms already
exist in the internal memory. Refer to steps 18 to 20.
Creating the Waveform
4.
Press the “
Edit...
” soft key of the GO/NO-GO top menu. The zone creation menu will
appear.
5.
Press the “
NEW...
” soft key and the base trace menu will appear.
6.
Press the soft key corresponding to the desired base trace. To edit the entire waveform,
continue with step 7; to edit a part of the waveform, continue with step 10.
For DL1520/DL1520L, menus such as “
TRACE1
”, “
TRACE2
” and “
MATH
” are
displayed instead of the base trace menu shown below.
Relevant Keys
Rotary knob
Soft keys
SELECT key
ESC key
<, > keys
GO/NO-GO key