9-12
IM 701510-01E
Points to note relating to parameter determination
• Trigger mode is automatically switched to single mode irrespective of the selected trigger
mode during GO/NO-GO determination. However, determination will be repeated if the
determination sequence has been set to “CONTINUE”. Even in case the GO/NO-GO
mode (at the GO/NO-GO top menu) is changed to OFF again, the setting at the trigger
mode selection will not become valid again.
• While both pre-zoom waveform and its expanded waveform are displayed at the same
time, the expanded waveform will become the object of determination.
• The following operations are not possible after having pressed the “EXEC” soft key. In
order to perform these operations, it is necessary to set “Mode” (at the GO/NO-GO top
menu) to “OFF”, after having pressed the “ABORT” soft key.
• Initialization of the settings
• Automated measurements
• Auto set-up
• Changing of expansion settings
• Cursor measurements
• Display of an X-Y waveform
• Action-on trigger
• During an action is being executed after GO/NO-GO determination, no waveform data
will be acquired.
• In case the linear scaling function has been set to ON, you can enter linear scaled values
at the GO/NO-GO parameter setting menu.
• Only waveforms displayed on screen are available for judgment. Acquisition-memory
data is not available.
9.2 Judging using Measured Values of Waveform Parameters