8-8
IM 701510-01E
Selecting the measurement items
The measurement items can be selected from 20 different ones. For the meaning of each
item, refer to the description on page 8-11.
4.
Press the “
Measure
” soft key at the automated measurement setting menu to display
the item selection menu.
5.
Press the soft key corresponding to the desired trace or loaded waveform number.
For DL1520/DL1520L, menus such as “
TRACE1
”, “
TRACE2
”, “
MATH
”, “
LOAD1
” and
“
LOAD2
” are displayed instead of the trace/load waveform selection menu shown below.
6.
Use the rotary knob to highlight the desired measurement item.
7.
Press the
SELECT key
to select an item. The selection mark in front of the item will be
highlighted to indicate it has been selected. Press the
SELECT key
again to undo this,
or to reset all selection at once, press the “
ALL CLEAR
” soft key.
8.
Repeat step 6 and 7 to select all the desired measurement items. The display of the
measurement items depends on the number of waveforms to be measured and the
selected items. Refer to page 8-13 for more details.
When selected, the mark
becomes .
Use the rotary knob to first
highlight an item, and then
(de)select it.
Switches between menus
For DL1520/DL1520L,
another menu is displayed.
Setting the distal/proximal values
The two levels which are used as the reference for the measurement of rise and fall times
can be set. For more details, refer to the description on page 8-11.
9.
Use the rotary knob to select either the “
PROX
” or “
DISTAL
” setting and press the
SELECT key
to display the value setting box.
10.
Use the rotary knob to set the desired value. Press the
RESET key
to restore the new
value to the default value.
11.
Press the
SELECT key
or the
ESC key
to close the value setting box.
12.
Repeat steps 9 to 11 to set either the “
PROX
” or “
DISTAL
” settings.
Value setting box;
set the value using the rotary knob.
13.
Repeat steps 5 to 12 in case you want to select measurement items for another trace or
loaded waveform. In case you want to select the same measurement items for all traces
and loaded waveforms, select “
Copy To All Trace
” using the rotary knob and press
the
SELECT key
.
Returning to the top menu
14.
Press the “
To Topmenu
” soft key or the
ESC key
to return to the automated
measurement setting menu.
8.3 Measuring Waveform Parameters Automatically