9-10
IM 701510-01E
Setting action after determination
18.
After having selected the “
Action
” setting using the rotary knob, press the
SELECT
key
to select the desired action. After a “NO-GO” determination, the data can be
recorded in the following manners.
“NONE”
: No action;
“PRINT”
: A hardcopy of the screen will be printed out on the built-in printer;
“FD(P-P)”
: All waveforms displayed on the screen will be saved to the floppy disk
as P-P compressed data;
“FD(ACQ)” : Acquisition data of all waveforms displayed on the screen will be saved
to the floppy disk as binary data.
19.
After having selected either “
FD(P-P)
” or “
FD(ACQ)
”, move to the
“
FileName
”setting and press the
SELECT key
. The keyboard will be displayed. After
having selected either “
NONE
” or “
”, continue with step 21.
20.
Use the keyboard to enter the name of the file. The entered filename will be displayed at
the “
FileName
”setting. For details how to operate the keyboard, refer to page 4-8.
Returning to the top menu
21.
Return to the GO/NO-GO top menu by pressing the
ESC key
.
Sets determination conditions for each parameter
(For DL1520/DL1520L, 1/2/Math is provided
instead of 1/2/3/4/ in Tracemenu.)
Selects logic
Selects sequence
Selects action after determination
Enters the file name using the keyboard
Executing the determination function
22.
Press the “
EXEC
” soft key. The name of the soft key will change to “
ABORT
”.
For the display which appears during measurement, refer to page 9-5.
Aborting the determination function
23.
Press the “
ABORT
” soft key. The name of the soft key will change to “
EXEC
”.
9.2 Judging using Measured Values of Waveform Parameters