Error
Messages,
Maintenance,
and
Inspection
11-7
IM 755601-01E
11
11.4 Adjustments
To maintain high measurement accuracy, we recommend that the instrument be
adjusted once a year.
Instruments to Prepare
Standard resistor
Nominal Value Accuracy
1
Ω
±
50 ppm
10
Ω
±
25 ppm
100
Ω
to 100 k
Ω
±
15ppm
1 M
Ω
±
25 ppm
10 M
Ω
±
50 ppm
100 M
Ω
±
100 ppm
Recommended instrument: 4808 (Fluke), 2792 (YOKOGAWA)
Environment and Conditions while making the adjustment
Ambient temperature: 23
±
1
°
C
Relative humidity:
45 to 75%RH
Supply voltage:
100 V
±
5%
Frequency:
(50/60 Hz)
±
1 Hz
Warm-up:
At least 2 hours for the standard resistor and 60 minutes for this
instrument before making the adjustment
Precautions while making adjustments
Connect the power cord of the instrument to a three-pole power outlet that has a
protective grounding terminal.