6-2
IM 755601-01E
6.1 Store/Recall Measured Data
Note
• Pressing the STORE DATA (SHIFT+0) key in the measurement mode while the store
operation is in progress (“STORE DATA” indicator is ON) aborts the operation and continues
the measurement.
• Pressing the RECALL DATA (SHIFT+ .) key while the recall operation is in progress
(“RECALL DATA” indicator is ON) aborts the operation and continues the measurement.
Explanation
Storing measured data
Up to 2000 measured data sets can be stored in the internal memory. The data are
cleared when the power is turned OFF or the settings are initialized. By using the
optional Centronics interface, statistics on the stored data can be computed and printed
to an external printer. For the procedures related to printing, see section 6.2.
Timing used to store or recall data
When the “STORE DATA” or “RECALL DATA” indicator is ON in the measurement
mode, measured data are stored or recalled every time a trigger occurs.
The timing used to store or recall data varies depending on the trigger mode as follows:
EXTERNAL: Data are stored or recalled when a trigger is activated using the input
signal at the external trigger input terminal or the number 8 pin of the
handler interface.
MANUAL:
Data are stored or recalled when the TRIG key is pressed or when a
trigger is activated using a communication command.
INTERNAL: Data are stored or recalled at measurement intervals according to the
specified measurement time. For details regarding the measurement
interval, see page 4-11.
Precautions to be taken when storing data
• The STORE DATA (SHIFT+0) key can be used even in the measurement mode
(when the measurement is in progress).
• Executing a store operation, clears the previously stored data.
• Data cannot be stored when the “RECALL DATA” indicator is ON. An error (831) is
displayed in this case.
• When the store operation is aborted, data stored up to that point can be recalled.
Precautions to be taken when recalling data
• Data are recalled in order from the first data in the memory. After recalling all of the
data, the instrument continues the measurement.
• Data cannot be recalled when the “STORE DATA” indicator is ON. An error (832) is
displayed in this case.
• If there are no stored data in the memory, an error (830) is displayed for
approximately one second, and the display returns to the top menu of the SETUP
mode.