7.1 Handler Interface Functions and Specifications
Output Signal Behavior
The behavior of the INDEX, EOM, NO CONTACT, HI, IN, and LO output signals varies
depending on the trigger mode as follows:
When the trigger mode is set to EXTERNAL
INDEX:
The signal is generated (set to L) when a trigger occurs, regardless of
the measurement result.
EOM:
The signal is generated (set to L) when the measurement completes
once, regardless of the measurement result.
NO CONTACT: The signal is generated (set to L) when the measurement results in a
contact check error (nC) or when the test current is abnormal (CF).
HI:
The signal is generated (set to L) when the comparator result of the
measured value is HI.
The signal is also generated (set to L) when the measurement results
in over-range (oL), contact check error (nC), or when the test current is
abnormal (CF).
IN:
The signal is generated (set to L) when the comparator result of the
measured value is IN.
LO:
The signal is generated (set to L) when the comparator result of the
measured value is LO.
When the trigger mode is set to MANUAL or INTERNAL
INDEX:
The signal is not generated (stays at H).
EOM:
The signal is not generated (stays at H).
NO CONTACT: The signal is generated (set to L) when the measurement results in a
contact check error (nC) or when the test current is abnormal (CF).
HI:
The signal is generated (set to L) when the comparator result of the
measured value is HI.
The signal is also generated (set to L) when the measurement results
in over-range (oL), contact check error (nC), or when the test current is
abnormal (CF).
IN:
The signal is generated (set to L) when the comparator result of the
measured value is IN.
LO:
The signal is generated (set to L) when the comparator result of the
measured value is LO.