Chapter 11 Error Messages, Maintenance, and Inspection
LOCAL key
GP-IB connector
11.1 Troubleshooting
Troubleshooting
• For corrective actions when a message appears on the screen, read the following pages.
• If servicing is required or if the instrument does not operate properly after taking the
following corrective actions, contact your nearest YOKOGAWA dealer.
Problem
Check Item
Reference Page
Nothing is displayed when the
Is the power cord securely connected to the power connector on
–
power switch is turned ON.
the instrument?
Is the power plug on the other end securely connected to the
power outlet?
Displayed data are odd.
Is there a possibility that noise is affecting the measurement?
–
Are the ambient temperature and humidity within the allowed
3-2
range as stated in the specifications?
Cannot make measurements.
Is the instrument in the measurement mode?
4-2
Keys do not operate.
Is the REMOTE indicator turned OFF?
9-1
Is the instrument in the SETUP mode?
4-2, 5-1
Is the HOLD signal of the handler interface inactive?
5-1, 7-1
The instrument cannot be
Are the communication specifications between the instrument
8-1
setup or controlled via the
and the controller properly matched?
serial interface.
The instrument cannot be
Does the GP-IP address assigned to the instrument match
9-6
setup or controlled via the
the address specified in the program?
GP-IB interface.
Are the electrical and mechanical specifications of the IEEE
9-2
St’d 488.1-1987 satisfied?
Cannot print
Is the printer in the proper state (on-line) to allow printing?
–