4-10
IM 755601-01E
Selecting the edge
6.
5.
4.
2.
3.
MISC
SHIFT
REF
LIMIT
k
Ω
REF
LIMIT
ENTER
Ω
,%
ENTER
Ω
,%
[Main display]
[Main display]
Manual trigger (only when the trigger mode is set to MANUAL)
1.
Press the
SHIFT
SETUP/MEAS
R/%
key to switch to the measurement mode.
2.
Trigger is activated every time the
TRIG
MODE
TRIG
key is pressed.
Explanation
CAUTION
Applying a voltage outside the 0 to 5 V range to the external trigger input
terminal (TRIGGER IN) can damage the instrument.
External trigger input terminal (TRIGGER IN)
Specifications
Connector type
: BNC
Input level
: CMOS
Minimum pulse width : 100
µ
s
Input circuit (non-isolated)
10 k
Ω
+5 V
+5 V
100
Ω
CMOS
4.6 Using the Trigger Function