![YOKOGAWA 7556 User Manual Download Page 89](http://html.mh-extra.com/html/yokogawa/7556/7556_user-manual_911164089.webp)
10-15
IM 755601-01E
Communication
Commands
10
10.2.2
CHECk(contact CHECk) Group
The commands in the CHECk(contact CHECk) Group deal with the contact check function.
These commands can be used to make the same settings and inquiries as when the CHECK key on the front panel is
pressed.
:CHECk
:
MODE
<Space>
OFF
BEFore
AFT
MIN
er
MAX
?
LEVel
<Space>
<Resistance>
?
?
;
:CHECk?
Function
Queries all settings related to the contact check
function.
Syntax
:CHECk?
Example
When contact check is OFF
:CHECK?
→
:CHECK:MODE OFF
When contact check is ON (BEFore)
:CHECK?
→
:CHECK:MODE BEFORE;LEVEL 5
:CHECk:LEVel
Function
Sets the contact check level or queries the
current setting.
Syntax
:CHECk:LEVel {<Resistance>|MAX|MIN}
:CHECk:LEVel?
<Resistance> = 1 to 30
Ω
(Resolution 1
Ω
)
MAX = 30
Ω
MIN = 1
Ω
Response
<NR1> form
Example
:CHECK:LEVEL 5OHM
:CHECK:LEVEL?
→
:CHECK:LEVEL 5
:CHECk[:MODE]
Function
Turns ON/OFF the contact check function and
sets the check timing or queries the current
setting.
Syntax
:CHECk[:MODE] {OFF|BEFore|AFTer}
:CHECk:MODE?
OFF = Contact check: OFF
BEFore = Contact check: ON, check before
measurement
AFTer = Contact check: ON, check after
measurement
Response
{OFF|BEFore|AFTer}
Example
:CHECK:MODE OFF
:CHECK:MODE?
→
:CHECK:MODE OFF
10.2 Commands