2. COMPONENT NAMES AND FUNCTIONS
Operation Principles
The operation principles of WG251/1001 for each process is defined as follows.
(1) Boiler drainage
●
Turn ON (|) ELB and press the POWER key.
⑩
Solenoid valve for boiler drainage
opens for about 30
seconds to perform initial boiler drainage (WG251 only). Boiler drainage is also performed once every
five hours of distillation.
(2) Boiler water supply and distilling operation
●
When boiler drainage completes,
④
Solenoid valve for raw water supply
and
⑰
Solenoid valve for
boiler water supply
open simultaneously. Then water is supplied to
⑫
Boiler.
When
⑧
Float switch (1)
in
the
⑦
Float cylinder
detects the water level,
⑬
Heater
is energized to start distillation. At this moment,
③
Solenoid valve for cooling water
opens as well. The water supply to
⑫
Boiler
is regulated by opening
and closing
④
Solenoid valve for raw water supply
and
⑰
Solenoid valve for boiler water supply
, using
⑨
Float switch (2)
.
To water
collection table
To water
collection table