23
3. PRE-OPERATION PROCEDURES
Installation Procedure
6. Connect ion-exchange resin cartridge (CPC-S) securely.
●
Connect ion-exchange resin cartridge (CPC-S) according to the following procedure.
Remove the rubber caps from the IN and OUT of the cartridge.
(1) Make sure that ELB on unit is "OFF (
○
)" and that water tap is closed.
(2) Take out ion-exchange resin cartridge from among the accessories.
(3) Install ion-exchange resin cartridge (CPC-S) on the mount in unit (Fig. 01).
(4) Secure ion-exchange resin cartridge with fixing band of the mount (Fig. 02).
(5) Push the one-touch joint marked with (IN CPC-S) into the inlet (on left) of ion-exchange resin
cartridge until it clicks.
Push the one-touch joint marked with (OUT CPC-S) into the outlet (on right) of ion-exchange
resin cartridge until it clicks (Fig. 03).
(6) Turn ON (|) ELB and wait for about 10 seconds. Press the reset switch on the back of front door
for two seconds (Fig. 04).
* One-touch joints may be stiff at the outset. Note that forcing one-touch joint diagonally into plug-in
port may result in damage to one-touch joint or plug-in port. If it is difficult to connect one-touch joint
to a cartridge, take the cartridge off the resin cylinder mounting plate, and carefully push the joint into
plug-in port.
* After inserting one-touch joint, pull it once to confirm that one-touch joint is securely connected, and
will not come off
* To remove the ion-exchange resin cartridge, pull the one-touch joint frontward with the black portion
pressed in. Be careful that water may drip from the cartridge when the one-touch joint is removed.
* Connecting the IN and OUT of each resin cartridge in reverse may cause malfunction. Use caution
when making connection of the cartridges.
IN
リセットスイッチ
IN
OUT
OUT
Fig. 01
Fig. 02
Fig. 03
Fig. 04
Reset switch