32
5. OPERATION PROCEDURES
Measured-amount Collection of Distilled Water
1
.
Switching water collection mode
Switch the collection mode lamp from
"CONTINUOUS" to "MEASURED" by pressing the
MODE key.
2. Setting the volume of water collection
Enter the volume of distilled water collection
The setting screen is common to ion exchanged
water and distilled water.
①
Enter the volume of water to be collected, using
the
▲▼
keys.
Setting range: up to 30 L (WG251), up to 100 L
(WG1001)
②
After setting, press the DISTILLED WATER key
to collect water.
3. Collecting distilled water
①
The value of collection volume counts down
while dispensing water in measured-amount
mode.
②
When water collection is completed, display
shows " Smpl Qty 0.0L".
③
Display returns to the screen in Section 2.
* Pressing the DISTILLED WATER key while
dispensing water stops operation and resets
collection volume.
* The pump may generate a loud noise when
dispensing water, but this is not a failure of unit. (If
the noise persists, remove membrane filter and
dispense distilled water to bleed air)
Distilling
Dstld Wtr Cndvty
0.85×10
-4
S/m
Smpl Qty 0.1L
Distilling
Dstld Wtr Cndvty
0.86×10
-4
S/m
Smpl Qty 26.0L
①
②
Smplng Dstld Wtr
Dstld Wtr Cndvty
0.85×10
-4
S/m
Smpl Qty 18.2L