52
8. EXTENDED STORAGE AND DISPOSAL
Extended Storage
WARNING
When unit will be out of service for an extended period.
Be sure to turn OFF (
○
) ELB and close water tap for safety. Leaving the water stored in boiler
and distilled water tank will allow the growth of bacteria or algae, impairing the water quality.
Discharge water by following the steps below.
Drain boiler
(1) When draining boiler, turn OFF (
○
) ELB,
close water tap and wait for at least 30
minutes to dissipate heat, and then open
front door and turn on boiler drain cock.
(2) Open boiler drain cock.
(3) Check that all the water in boiler and float
cylinder has been discharged.
(4) Be sure to close boiler drain cock. If boiler
drain cock is open at the next use, all the
water will be discharged from drain cock
and will not flow into boiler, hindering
distillation.
When unit will be out of service for an extended period.
●
Turn OFF (
○
) ELB on the right side of unit.
●
Be sure to close the tap.
●
If unit is used in a place where it becomes extremely cold in winter, beware of freezing of
tank, boiler, and condenser while unit is in storage.
Disposal Considerations
CAUTION
Unit disposal
●
Dispose of this unit in accordance with local laws and regulations.
●
Do not leave unit where it may be unattended, or in a location where children may have
access.
Dispose of or recycle this unit in a responsible and environmentally friendly manner.
Yamato Scientific Co., Ltd. strongly recommends disassembling unit, as far as is possible, in order to
separate parts and recycle them in contribution to preserving the global environment. Major
components and materials, comprising WG series unit are listed in the table below
CLOSE