47
7. INSPECTION AND MAINTENANCE
Washing of Distiller
●
Removing distiller
1. Turn OFF (
○
) ELB.
2. Close the water tap.
3. Turn OFF (
○
) ELB and wait for at least 30 minutes to dissipate heat, and then open front door and
turn on boiler drain cock.
4. Remove the hose connected to
①
boiler
and
②
condenser
. Remove the distilled water outlet and
boiler water supply/drain port by twisting the hose band with a tool to disengage the meshing
portion (serrated portion). Remove the glass carefully, the glass may break if excessive force is
applied.
5. Remove hole plugs, which cover screw holes, from the left side panel, remove four screws with a
Phillips screwdriver, and remove the left side panel.
6. A terminal block will appear in the upper right corner of the frame. Loosen four screws on the right
side of the terminal block, and pull out the heater lead wire terminals. (since WG1001 employs two
heaters, there are eight screws)