A P P E N D I X E
E-2
COMPUTER UPDATE PROCEDURE
The following items are required to perform a instrument firmware update
using a PC serial port.
•
Computer with an available serial port
•
Instrument with AC adapter
•
Update firmware disk
•
Interface cable (P/N SE108-21-01) and Null Modem adapter
(P/N SE108-DB25FN-01)
NOTE: Computer screen savers must be disabled during
the update procedure. Once the update has been started,
the old firmware version is permanently erased.
Step 1 Cabling Connection Procedure
•
Insert the modular end of the interface cable into the I/O port on the instrument.
•
Insert the DB25 adapter end of the interface cable into one end of the null
modem adapter.
•
Attach the null modem adapter to an available “RS-232” I/O port on the
computer.
Step 2 Setting Instrument to Receive Update
•
Apply AC power to the instrument.
NOTE:
Make sure the modem baud is set to "57.6K." Press
netwk
key on p5
menu. Press
cfg
key,
mod
key, and
baud
key. Edit baud if required. Refer to
Section Two in the 891/892 Networking Installation and Operation Manual for
additional information.
•
Enter the
Network Menu
by pressing key
ll
(netwk) on page 5 (p5)
Function
Menu
.
Instrument
I/O Serial Port
Modular End
Interface Cable P/N SE108-21-01
DB25 Adapter
Null Modem
P/N SE108-DB25FN-01
To RS-232
Port
Summary of Contents for 891
Page 1: ...891 892 A U T O M A T I C S T R I P R E A D I N G D E N S I T O M E T E R S Operation Manual...
Page 2: ......
Page 4: ......
Page 10: ...I N T R O D U C T I O N vi...
Page 34: ...S E C T I O N T H R E E 3 8...
Page 46: ...S E C T I O N F O U R 4 12...
Page 52: ...S E C T I O N F I V E 5 6...
Page 58: ...S E C T I O N S I X 6 6...
Page 60: ...S E C T I O N S E V E N 7 2...
Page 66: ...S E C T I O N E I G H T 8 6...
Page 68: ...A P P E N D I X A A 2...
Page 73: ...D 1 A P P E N D I X D Parts List and Packaging Drawings Parts List...
Page 74: ...A P P E N D I X D D 2 Packaging Drawing...
Page 78: ...A P P E N D I X E E 4...
Page 79: ......