I N T E R N E G A T I V E O P E R A T I O N ( 8 8 2 O N L Y )
6-3
VIEWING INTERNEGATIVE DATA
The View function allows you to view the density difference values and
recommended filter pack changes. Individual internegative steps can also be
viewed.
Density Difference
Density difference values are calculated by simply subtracting the density
values of the higher numbered step from the densities of the lower numbered
step. These values are used to look-up your filter pack changes from the table.
1. Press
key
l
(view).
STEP PAIR= 3, 13
view +step– auto
2. Press
key
lll
(
∆
fp) to display the filter pack changes from the internal table.
These numbers are changed (indicated by the Greek letter
∆
) and should be
added to the filter pack used for exposing this internegative.
D= 0.62 0.75 0.80
exit
∆
fp
3. Press
key
lll
(
∆
den) to redisplay the density difference values. Press
key
llll
(exit) to return to the Step Pair menu to allow further viewing or
calculations.
∆
FP —03C +05M +02Y
exit
∆
den
Individual Steps
1. Press
key
ll
(
+
) to increment step or key
lll
(
−
) to decrement step. Press
key
llll
(exit) to return to the Step Pair menu to allow further viewing or
calculations.
0.20 0.47 0.83
#21 +step– exit
Summary of Contents for 891
Page 1: ...891 892 A U T O M A T I C S T R I P R E A D I N G D E N S I T O M E T E R S Operation Manual...
Page 2: ......
Page 4: ......
Page 10: ...I N T R O D U C T I O N vi...
Page 34: ...S E C T I O N T H R E E 3 8...
Page 46: ...S E C T I O N F O U R 4 12...
Page 52: ...S E C T I O N F I V E 5 6...
Page 58: ...S E C T I O N S I X 6 6...
Page 60: ...S E C T I O N S E V E N 7 2...
Page 66: ...S E C T I O N E I G H T 8 6...
Page 68: ...A P P E N D I X A A 2...
Page 73: ...D 1 A P P E N D I X D Parts List and Packaging Drawings Parts List...
Page 74: ...A P P E N D I X D D 2 Packaging Drawing...
Page 78: ...A P P E N D I X E E 4...
Page 79: ......