S E C T I O N F O U R
4-8
AIM VALUE ADJUSTMENTS (AVA)
Certain conditions may require value adjustment of your reference strips. The
aim value adjustments (AVA) are range limited and will not exceed the aim
value adjustment tolerances (AVAT’s) set in the instrument.
Simultaneously pressing key
lll
and key
llll
labeled MENU at anytime aborts
the AVA function. Any adjustments that had been made prior to exiting are
not saved.
1. Press
key
llll
(edit) located on Function Menu page (p5) to enter Editor
Menu screen.
FUNCTION MENU
p5 netwk cal edit
2. Press
key
l
(ava) located on Editor Menu screen to enter Select Channel
screen.
EDITOR MENU
ava cnfg del time
3. Press
key
l
(chan) to select reference channel. Press key
llll
(go>) to
display AVA edit screen.
SELECT CHANNEL
pap
go
>
NOTE:
NO REF INSTALLED
appears in the display if no reference is
available in the selected channel.
4. Adjust aim values for all required fields. Press key
l
(D-min, etc.) to page
through available fields. Note, press the key
l
again after the last field
displays to advance the screen to the Aims Adjusted screen.
•
Press key
llll
(
→
) to move the cursor to desired location. Press key
ll
(
↑
)
to increase value and key
lll
(
↓
) to decrease value.
0.00 0.00 0.00
D—min
↑
↓
→
—
Cursor
Summary of Contents for 891
Page 1: ...891 892 A U T O M A T I C S T R I P R E A D I N G D E N S I T O M E T E R S Operation Manual...
Page 2: ......
Page 4: ......
Page 10: ...I N T R O D U C T I O N vi...
Page 34: ...S E C T I O N T H R E E 3 8...
Page 46: ...S E C T I O N F O U R 4 12...
Page 52: ...S E C T I O N F I V E 5 6...
Page 58: ...S E C T I O N S I X 6 6...
Page 60: ...S E C T I O N S E V E N 7 2...
Page 66: ...S E C T I O N E I G H T 8 6...
Page 68: ...A P P E N D I X A A 2...
Page 73: ...D 1 A P P E N D I X D Parts List and Packaging Drawings Parts List...
Page 74: ...A P P E N D I X D D 2 Packaging Drawing...
Page 78: ...A P P E N D I X E E 4...
Page 79: ......