94
ba31116e05
07/2014
What to do if...
OxiTop
®
Control OC 110
Which measuring head
belongs to which mea-
surement curve?
Prepare the relation separate curve – number of measuring head; see sec-
tion 5.5 E
Read number of measuring head out of headline
Have measuring head shown; see section 5.4.1 S
The precision and sealing
of the system sample bot-
tles plus measuring head
should be checked
Measuring head check
(single check)
(To which running sam-
ple does the selected
measuring head belong?)
Single check method:
Menu
GLP/TOOLS
- Check
-
info
The controller guides you further through the operation.
All single information on the state of the measuring heads is listed, i.e. the
measuring head responds.
The check can be undertaken at any time without affecting the measure-
ment that is running.
If no reaction at all can be determined, the batteries should be changed and
the measuring head check repeated.
The measuring precision cannot be checked with this!
The measuring precision
of the measuring head
should be checked.
Perform measuring head
reaction test
Menu
GLP/TOOLS
- Check
- All
:
All optically attainable measuring heads must flash for approx. 5 seconds
independent of their status.
The check can be undertaken at any time without affecting the measure-
ment that is running.
This test only checks the reaction to commands.
If no measuring head reacts, the controller should be checked.
Cause
Remedy
Summary of Contents for OxiTop Control OC 110
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Page 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Page 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Page 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Page 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Page 101: ......