OxiTop
®
Control OC 110
Operating mode: Pressure p
ba31116e05
07/2014
45
7
Operating mode:
Pressure p
Switch to the
Pressure p
operating mode: see section 9.4.1 O
7.1
Hazards during the determination of biogas
CAUTION
The pressure sensor in the measuring head may be destroyed by
a pressure of 2000 hPa or higher.
CAUTION
During the anaerobic degradation of organic material in sealed
equipment, high pressure can develop due to biogas (main
components: methane and carbon dioxide).
The following hazards can occur:
Risk of bursting due to the high pressure and risk of injury
caused by flying glass splinters.
Risk of explosion due to development of flammable gas
mixtures if high concentrations of methane escape.
Take the following precautionary measures for safe operation:
Only use WTW measurement bottles of the MF/... series,
which are especially designed for biogas determination
(see WTW catalog).
Only use measurement bottles that are not damaged (e.g.
are not cracked, chipped or scratched).
Determine the reaction conditions (temperature, quantity)
under which no dangerous excess pressure develops by
means of monitored preliminary tests (e.g. with diluted ma-
terial).
For unmonitored measurements, include a buffer for the de-
veloping pressure, as the composition of biological materi-
al can considerably vary. Therefore, a pressure that is
higher than planned on can develop.
Only use the corrosion resistant OxiTop
®
-C/B measuring
heads. They were designed especially for biogas applica-
tions.
Please not that the OxiTop
®
measuring head displays a
pressure of up to 1350 hPa max.If the pressure is greater, -
1000 hPa is displayed.
Summary of Contents for OxiTop Control OC 110
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Page 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Page 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Page 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Page 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Page 101: ......