Operating mode: BOD Special
OxiTop
®
Control OC 110
44
ba31116e05
07/2014
The further steps from starting the samples to evaluation are as
described for
BOD Standard
operating mode (see section
Additionally displayed information:
Type of the measurement, run duration (setting see section
9 GLP/TOOLS, page 57), measuring range, end date, Id
number.
GLP
display => the
GLP
function (see section 9 GLP/
TOOLS, page 57) is switched on.
Possibly
Temp
display => The
AutoTemp
function (see sec-
tion 9 GLP/TOOLS, page 57) is switched on (only with run
durations > 1 day).
Change the ID number for the additional identification of the
sample (e.g. sampling location) as follows:
Move the cursor to Id number with
<
><
>
.
Confirm with
<RUN/ENTER>
.
Set the required Id number (setting range 001 ... 255) with
<
><
>
.
Confirm with
<RUN/ENTER>
.
Print all the sample information with
<PRINT>
Confirm the start of measure-
ment. Contact selection:
RUN/ENTER
The calculation of the BOD value in the OC 110 controller is based
on the assumption that the D. O. partial pressure in the bottle is at
least 60 hPa after the measurement is over.