GLP/TOOLS
OxiTop
®
Control OC 110
72
ba31116e05
07/2014
Evaluation of the Cal
test
CHECK
menu.
Use
<
><
>
to select the
- Cal
Test
menu item.
The instrument automatically
allocates the sample number (in
the header line).
The filling volume (164 ml) and
the type together with the run
time of 5 days are preset.
For information on the further handling of the sample up to the
“finished” status: See section 5.2 S
The Cal sample appears together with the other samples in the
SAMPLE STATUS TYPE
list (
sample management). The BOD
type is
CAL
:
SAMPLE STATUS TYPE
list
(sample management)
RUN/ENTER
Use
<
><
>
to select the test
sample.
Entry into the evaluation.
Display of the selected sample
as a curve with measurement
value data.
Compare measurement value
with lot test value (according to
operating manual of OxiTop
®
PM) .