Operating mode: BOD Standard / BOD Routine
OxiTop
®
Control OC 110
26
ba31116e05
07/2014
5.3
Calling up all data
This function is used to call up the data of all measuring heads, no
matter in which operating mode they were started.
To call up the data of individual measuring heads: see section 5.4
SAMPLE STATUS TYPE
SELECT ACTION
menu (com-
munication with the measuring
heads):
Preselected:
- Start sample
.
Select the
- Call up all data
menu
item with
<
><
>
Point the controller at the measuring
heads:
Queries the data of all active
measuring heads in the scanner
mode.
The controller stores the data
obtained and updates the
SAMPLE STATUS TYPE
list
(
sample management).
Duration of a pass: approx. 3
seconds for 12 measuring heads
in a stirring system.
40 cm to 1 m
RUN/ENTER
If the measuring heads do not all respond in the 1st pass of the data query,
the controller searches for the missing measuring heads in the scanner mode
for approximately a further 7 seconds. Without having to press another key,
you can continue the data queries at other locations (e.g. other stirring plat-
forms or other thermal cabinets).
Approximately 7 seconds after the last call up was performed, the scanner
mode breaks off automatically and the following display appears:
Summary of Contents for OxiTop Control OC 110
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Page 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Page 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Page 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Page 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Page 101: ......