OxiTop
®
Control OC 110
Operating mode: BOD Special
ba31116e05
07/2014
41
6
Operating mode:
BOD Special
To change to the operating mode
BOD Special
6.1
Sample preparation
See application reports.
Screw the OxiTop
®
measuring heads onto the measuring vessels
and close them tightly.
NOTE
Never use joint grease or other lubricants for your OxiTop
®
measuring heads. Some of these products contain solvents that
can cause severe damage to the plastic housing of the measuring
head.
The sealing of the measuring vessels is perfectly adequate
without grease. However, you should always wipe off heavy
contamination and particles on the sealing surfaces of the rubber
sleeves and OxiTop
®
measuring heads. WTW accepts no liability
for damage due to the use of joint grease.
6.2
Starting the measurement
Before starting the measurement, set the temperature and air pressure
(see section 9 GLP/TOOLS, page 57).
In the
BOD Special
mode there are 6 variable parameters:
SELECT ACTION
menu (com-
munication with the measuring
heads):
Preselected:
- Start sample
.
Parameters
Default setting Admissible setting range
Bottle volume
510 ml
50 to 9.999 ml
Filling volume
164 ml
10 ml to bottle volume minus 10 ml
Dilution
1 + 0
1 + 0 to 1 + 99
Measuring range
(BOD)
395 mg/l
up to 400,000 mg/l
Temperature
20°C
5 °C to 40 °C
(setting: see section 9 GLP/TOOLS,
page 57)
Air pressure
954 hPa
500 hPa to 1100 hPa (setting: see
section 9 GLP/TOOLS, page 57)
Summary of Contents for OxiTop Control OC 110
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Page 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Page 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Page 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Page 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Page 101: ......