OxiTop
®
Control OC 110
GLP/TOOLS
ba31116e05
07/2014
71
9.5.3
Controller information
9.5.4
cal-test
The
cal-test
function is used to test the sealing of the system measuring
head - rubber sleeve - sample bottle and the operability of the system
OxiTop
®
Control.
In the
BOD Standard
mode with the
GLP
On
setting, the controller indi-
cates when the next cal test is due, depending on the set calibration
interval (see section 9.4.4 GLP (
To perform the test, you need the WTW test resource, OxiTop
®
PM,
order number 209 333.
Sample preparation: see operating manual of the OxiTop
®
PM test
resource.
Query stopped!
You can select between
- Continue query
(see above)
and
- Stop
(return to the
CHECK
menu)
CHECK
menu.
Use
<
><
>
to select the
-
Controller Info
menu item.
The display shows the following
information:
Number of measuring heads
for which there is still memory
capacity
Status of the supply batteries
Status of the data backup
batteries (
OK
/
LOBAT
).
Use
<
><
>
to scroll to the
specification of the
Software version
Series number
(can differ from the example
shown here)
Return to the
CHECK
menu:
With
<RUN/ENTER>
.
RUN/ENTER
Summary of Contents for OxiTop Control OC 110
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Page 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Page 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Page 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Page 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Page 101: ......