70
WIKA operating instructions pneumatic high pressure controller, model CPC7000
0019104001D 10/2018EN
EN
[:LEVel]
[:IMMediate]
[:AMPLitude] <n>
Sets the setpoint
[:AMPLitude]?
Returns the setpoint
:SLEW <n>
Sets the rate setpoint
:SLEW?
Returns the rate setpoint
:TOLerance?
Returns the stable window
:TOLerance<n>
Sets the stable window
CALCulate
:LIMit
:LOWer<n>
Set the minimum control limit
:LOWer?
Set the minimum control limit
:UPPer<n>
Set the maximum control limit
:UPPer?
Set the maximum control limit
:SYSTem
:DETECT SLOW | FAST | CANCEL
Not used, kept for backwards compatibility
:DETECT?
Returns “0”
7.11.4 SCPI Commands Error Messages and Error Codes
All remote commands sent to the CPC7000 are shown in the Trouble Shooting App, under the “Remote” button. If there is a syntax
error, an error message will appear below the errant command. Local errors and remote command errors will appear under the
“Error” button. Maximum of 100 errors are stored and can be retrieved.
7. Remote Operation