6.4.3 Control Settings Application
The Control Settings App allows the user to select and configure the control parameters for the instruments pressure
regulating module (PVR module). Figure 6.4.3 shows the Control Settings Application screen.
Figure 6.4.3 Control Settings App
6.4.3.1
Control Limits
The Maximum and Minimum Limit buttons in the Control Settings App provide a place to limit the set point value that can be
chosen in the Home App. These limits can only be set within the range of the active transducer. When the instrument is in
Autorange the limits can only be set within the range of the primary transducer which, by convention, will have the widest range.
The minimum limit must be lower than the maximum limit. The user cannot enter set points and thereby not control to pressures
outside of these limits. Figure 6.4.3.1 shows the data entry keyboard to enter the limits. Note that the keyboard displays the
maximum value that can be entered.
Figure 6.4.3.1Control limit setting