64
WIKA operating instructions pneumatic high pressure controller, model CPC7000
0019104001D 10/2018EN
EN
Step?
<sp>+n.nn<cr><lf>
Returns the control step for the instrument
Step%
Value in % of current range
Sets the control step in % of current range
Steppct
Value in % of current range
Sets the control step in % of current range
Steppct?
<sp>n.nn<cr><lf>
Returns the current step in % of current range
Switch_Start
Changes the screen to the Switch Test if on another
screen, begins the switch test with current parameters.
Switch_Stop
Stops the test and vents the system.
Switch_State?
<sp>(IDLE or RUNNING or COMPLETE)<cr><lf>
Returns state of the switch test
Switch_Low
Value inside upper and lower limits
Starting setpt for the switch test to begin at
Switch_Low?
<sp>+n.nn<cr><lf>
Returns starting setpt
Switch_High
Value inside upper and lower limits
Finishing setpt for the switch test to end at
Switch_High?
<sp>+n.nn<cr><lf>
Returns ending setpt
Switch_Fast
Value inside upper and lower limits
Rate at which switch test will go to the low setpt
Switch_Fast?
<sp>+n.nn<cr><lf>
Returns fast rate setpt
Switch_Slow
Value inside upper and lower limits
Rate at which switch test will go to the high setpt
Switch_Slow?
<sp>+n.nn<cr><lf>
Returns slow rate setpt
Switch_1_Up?
<sp>+n.nn<cr><lf>
Returns the pressure at which switch 1 tripped while
going up scale. Returns 0 if switch was not triggered.
Switch_1_Dn?
<sp>+n.nn<cr><lf>
Returns the pressure at which switch 1 tripped while
going down scale. Returns 0 if switch was not trig-
gered.
Switch_2_Up?
<sp>+n.nn<cr><lf>
Returns the pressure at which switch 2 tripped while
going up scale. Returns 0 if switch was not triggered.
Switch_2_Dn?
<sp>+n.nn<cr><lf>
Returns the pressure at which switch 2 tripped while
going down scale. Returns 0 if switch was not trig-
gered.
Switch_3_Up?
<sp>+n.nn<cr><lf>
Returns the pressure at which switch 3 tripped while
going up scale. Returns 0 if switch was not triggered.
Switch_3_Dn?
<sp>+n.nn<cr><lf>
Returns the pressure at which switch 3 tripped while
going down scale.
Tare
ON / OFF
Tares the reading to zero
Tare?
<sp> n.nn <cr><lf>
Returns value of Tare
Termchar
CCCC
Set the ouptput termination character(s). CR, LF,
CRLF, DEFAULT, EOI, NONE
Termchar?
Returns the termination character setting
Transfer_factory_to_linearity
Copy factory linearity coefficients to customer
Units
Units code or text in table below
Sets the instrument engineering units
Units?
<sp>CCCC<cr><lf>
Returns the instrument units in a text string
Unitbase1
Units code or text in table below
Sets the user 1 base engineering units
Unitbase1?
<sp>CCCC<cr><lf>
Returns the user 1 base units in a text string
Unitbase2
Units code or text in table below
Sets the user 2 base engineering units
Unitbase2?
<sp>CCCC<cr><lf>
Returns the user 2 base units in a text string
Unitfact1
Multiplier factor
Sets the user 1 unit mulitplier
Unitfact1?
<sp>n.nn<cr><lf>
Returns the user 1 unit multiplier
Unitfact2
Multiplier factor
Sets the user 2 unit mulitplier
Unitfact2?
<sp>n.nn<cr><lf>
Returns the user 2 unit multiplier
7. Remote Operation