61
WIKA operating instructions pneumatic high pressure controller, model CPC7000
0019104001D 10/2018 EN
EN
Errorno?
<sp>Enn-text<cr><lf>
Returns pcs400 error code and text
Filter
OFF, LOW, NORMAL, HIGH
Sets the reading filter 0, 80%, 92%, 95%
Filter?
<sp> (filter)<cr><lf>
Returns the reading filter.
Gasdensity
Value in lb/cuft, or “NITROGEN” or “DRYAIR”
Sets the head pressure gas density in lb/cuft
Gasdensity?
<sp>+n.nn<cr><lf>
Gets the head pressure gas density in lb/cuft
Gastemp
Value in degrees F
Sets the head pressure gas temperature
Gastemp?
<sp>+n.nn<cr><lf>
Gets the head pressure gas temperature
Gateway
nnn.nnn.nnn.nnn
Sets the Ethernet gateway address
Gateway?
<sp>nnn.nnn.nnn.nnn<cr><lf>
Gets the Ethernet gateway address
Height
Value in inches
Sets the head pressure height in inches
Height?
<sp>+n.nn<cr><lf>
Gets the head pressure height in inches
Id?
<sp>MENSOR,CPC7000, ssssss,v.v.vv<cr><lf>
Ssssss is the serial number,v.v.vv is the CPC7000
software version.
IP
nnn.nnn.nnn.nnn
Sets the IP address of the instrument
IP?
<sp>nnn.nnn.nnn.nnn<cr><lf>
Returns the IP address of the instrument
Keylock
YES or NO
Locks or unlocks the entire touch screen
Keylock?
<sp>(YES or NO)<cr><lf>
Returns Yes or No
Leak_Start
Changes the screen to the Leak Test if on another
screen, begins the leak test with current parameters.
Leak_Stop
Stops the test and vents the system.
Leak_State?
<sp>(IDLE or RUNNING or COMPLETE)<cr><lf>
Returns state of the leak test
Leak_Setpt
Value inside upper and lower limits
Starting point for the leak test to begin at
Leak_Setpt?
<sp>n.nn<cr><lf>
Returns starting setpt
Leak_Dwell
0 to 3600
The duration of the test during which a leak is mea-
sured
Leak_Dwell?
<sp>n.nn<cr><lf>
Returns the dwell time in seconds
Leak_Pre_Dwell
0 to 3600
Pre-dwell parameter is the time in between going to
measure mode, and the dwell time. Useful if there are
pressure transients after switching the measure.
Leak_Pre_Dwell?
<sp>n.nn<cr><lf>
Returns the pre-dwell time.
Leak_Delta_Limit
Value within the range of the sensor
Threshold for a passing/failing leak test.
Leak_Delta_Limit?
<sp>n.nn<cr><lf>
Returns the maximum pressure change allowed before
failure
Leak_Pass?
<sp>(YES or NO)<cr><lf>
Leak_Delta?
<sp>n.nn<cr><lf>
Returns the change in pressure since start of test.
List?
<sp>Pri,1,2;Sec,1,2;Ter,1,2;Bar,1<cr><lf>
Returns list of available transducers and turn downs for
the instrument
Listcal?
<sp>PRI,{sn},1,{mmddyyyy},2,{mmdd
yyyy};SEC,{sn},1,{mmddyyyy},2,{mmddyyyy};
TER,{sn},1,{mmddyyyy},2,{mmddyyyy} <cr>
<lf>
Returns the serial number of each installed transducer
and calibration dates for each range
Listrange?
<sp>PRI,1,min,max,2,min,max;SEC,1, min,max-
,2,min,max; TER,1,min,max,2,min,max <cr><lf>
Returns the ranges of the installed transducers.
Listsensorinfo?
<sp>PRI,SN,PTYPE,ACCURACY,1,min,max,2,min,-
max;SEC,SN,PTYPE,ACCURACY,1,min,max,2,min,-
max; TER,SN,PTYPE,ACCURACY,1,min,max-
,2,min,max
Returns the serial number, pressure type, accuracy,
and ranges of the installed transducers
Localgravity
Value in ft/s^2
Sets the local gravity in feet/sec^2