Message
Possible cause
Suggested action
Temperature not functional. 03CB00 Unable to send message
to the Temperature
module
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Temperature not functional.
03CC00
Temperature module
communication times
out
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Temperature not functional.
03CD00
Failed to upgrade the
temperature module
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Temperature not functional. 03CE00 Unable to read PIM file
Retry the device update.
Temperature not functional. 03CE01 Upgrade file directory
could not be accessed
Retry the device update
Direct mode reading timed out
Direct mode reading
times out
Direction mode reading times out
Tissue contact lost.
Lost tissue contact while
attempting to acquire a
temperature
measurement or
acquired measurement
was performed with
limited tissue contact.
Check the tissue contact and try the
measurement again.
Temperature module reset. 03D000 Temperature sensor reset
unexpectedly
None
Patient data messages
Message
Possible cause
Suggested action
Database schema out of
data; recreating.
The database was cleared due to a
schema update
Information status message; press OK button
to dismiss.
Database is unreadable
during startup;
recreating. 1F0001
The database was unreadable
during startup
Press OK button to dismiss.
Error accessing PDM
database; restarting PDM.
1F0002
Database corrupted when device is
in operation
Press OK button to dismiss.
Maximum number of
patient r Oldest
record overwritten.
Data was deleted as it contained
more than 50 records
Information status message; press OK button
to dismiss.
No data saved.
A manual save is not allowed
Information status message; press OK button
to dismiss.
Save successful.
A manual record was saved
Information status message; press OK button
to dismiss.
Service manual
Appendix 117
Summary of Contents for Spot Vital Signs 4400
Page 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Page 8: ...4 Symbols and definitions...
Page 34: ...30 Power up sequence...
Page 68: ...64 Field replaceable units...
Page 127: ...Interconnect diagram Service manual Appendix 123...
Page 128: ...124 Appendix...
Page 132: ......