Message
Possible cause
Suggested action
inspected by a qualified service
technician.
NIBP not functional. 050607
NIBP cannot clear the failsafe errors Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050608
NIBP safety processor has stopped
responding
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Stat mode requested too soon.
Clear to retry. 050609
NIBP excessive stat mode time. The
time between readings is less than
one minute and the readings plus
time between readings result in the
device taking over 15 minutes to
complete the averaging cycle.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Unable to determine NIBP; check
connections and tubing for kinks.
05060A
NIBP transducers are not matched The transducer are above 5 mmHg
and the difference in pressure is
greater than 40 mmHg. Check cuff
for pinched or obstructed tubing. If
the problem persists, have the
device inspected by a qualified
service technician.
NIBP not calibrated. Calibrate the
module. 05060B
Factory EEPROM checksum error on
NIBP. Units internal configuration
was corrupted
Calibrate the NIBP module. If the
problem persists, have the device
inspected by a qualified service
technician.
NIBP not functional. 05060C
NIBP command not implemented Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 05060D
NIBP wrong data count
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 05060E
NIBP data range error
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 05060F
NIBP no POST error to clear
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050610
NIBP cannot clear this POST error
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050611
NIBP command not command type Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050612
NIBP communication timeout
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician..
108 Appendix
Summary of Contents for Spot Vital Signs 4400
Page 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Page 8: ...4 Symbols and definitions...
Page 34: ...30 Power up sequence...
Page 68: ...64 Field replaceable units...
Page 127: ...Interconnect diagram Service manual Appendix 123...
Page 128: ...124 Appendix...
Page 132: ......