Attach the bezel to the front housing
Kit items:
•
107183, SRV KIT, LCD BEZEL
•
107213, SRV KIT, FRONT HOUSING W/ POWER SWITCH
1. Lay the bezel over the 6 bosses on the front housing, and thread the power switch flex cable
through the slot in the bezel.
2. Press the bezel down onto the 6 bosses on the front housing.
Remove the SureTemp probe end cap
On the inside of the back housing, remove the end cap by gently pulling outward.
Reassembly notes
Kit item:
•
107354, SRV KIT, PROBE BOX CAP
NOTE
Reassembly is the reverse of disassembly.
Remove the power switch
1. On the back of the front housing, carefully remove the Gaff tape placed over the power switch
flex cable.
Service manual
Disassembly and repair 57
Summary of Contents for Spot Vital Signs 4400
Page 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Page 8: ...4 Symbols and definitions...
Page 34: ...30 Power up sequence...
Page 68: ...64 Field replaceable units...
Page 127: ...Interconnect diagram Service manual Appendix 123...
Page 128: ...124 Appendix...
Page 132: ......