In addition, because the software upgrade is already loaded on your device, you can use the
rollback capability at a later time to implement the upgrade and complete the associated setup
activities. Follow the same steps presented above to activate the new firmware.
Upgrade the device firmware
Updating a device requires a PC, a USB cable and an Internet connection.
1. To launch the Welch Allyn Service Tool, from the Start menu select
All Programs
>
Welch
Allyn
>
Welch Allyn Service Tool
.
2. Click
Service.
3. Click
Login.
a. Enter the user name:
ADMIN
.
b. Leave the password blank.
4. Connect the device to the computer with a USB cable.
5. Turn on the device.
6. In the Welch Allyn Service Tool, under the Device List tab, click on the device and click
Select
.
7. Under the Welch Allyn Connex Device, select
Upgrade
.
The list of available upgrades appears.
8. Select the desired upgrade, click
Upgrade
and then click
Yes
.
When the upgrade is complete, the Upgrade tab will display
No Upgrade Available
.
9. Close the Device tab, and then close the Service Tool.
10. Disconnect the USB cable from the device, and then turn off the device.
Partners in Care service and support agreements
Comprehensive support agreements
Material no.
Item
S1-4400
4400, Comprehensive partnership program point of sale
S1-4400-2
4400, Comprehensive partnership program 1 year renewal
Service and repair training
NOTE
Required to be eligible to receive the Welch Allyn Service Tool, Gold edition.
Material no.
Item
4400REP-TRN
4400, Repair training
4400REP-WEB-TRN
4400, Repair web training
Service manual
Options and upgrades 85
Summary of Contents for Spot Vital Signs 4400
Page 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Page 8: ...4 Symbols and definitions...
Page 34: ...30 Power up sequence...
Page 68: ...64 Field replaceable units...
Page 127: ...Interconnect diagram Service manual Appendix 123...
Page 128: ...124 Appendix...
Page 132: ......