Material no. Description
Qty
Component
web
download
Welch Allyn Service Tool
welchallyn.com/en/service-support/service-
1
NIBP, Software updates and
calibration
4500-30
Blood pressure hose, 5 ft.
1
NIBP
620216
Fitting "Y" 1/8 X 1/8 X 1/8
1
NIBP
N/A
USB Flash Drive
1
Host Controller, Log File
Capture
Initialize the NIBP board
Functional verification and calibration must be performed each time the device is disassembled or
more frequently if your facilty's procedures require a full calibration on a specified basis. If you have
just replaced a Main PCA board, your first step before performing a full verification and calibration is
to initialize the NIBP Board.
1. Set up the equipment for full functional verification as shown earlier in this section.
2. Log in to the Welch Allyn Service Tool, highlight your device in the Device list, and click
Select.
This action opens the
Spot 4400
>
Device information
subtab. Depending on your
configuration, it could take 30-45 seconds to populate this subtab.
3. Select the
Verify and calibrate
subtab.
4. Select
NIBP sensor
in the Device field.
5. Click
Calibrate
.
This opens the Verify and calibrate NIBP Sensor dialog.
6. Click
Begin
.
A list of tests appears in the left pane of the dialog along with three controls in the right pane.
7. Click
Initialize NIBP Board
in the right pane.
A confirmation window appears with the message, "Are you sure you want to reset the
calibration offsets of the NIBP Board?"
8. Click
Yes
.
This action places two tests in the queue for initialization: the Power Calibration Test – Primary/
Safety and the Pressure Calibration – Primary/Safety.
NOTE
A message appears in the right pane to indicate the first
test is in progress. If the test passes, a green check appears to the
left of the test name, and the second test begins. If the test fails, a
red check appears to the left of the test name, followed by a
prompt to rerun the test. You cannot proceed to the second test
until the first test passes.
9. Rerun the Power Calibration Test as needed until the test passes.
10. If you did not set up your equipment for a full calibration intially, connect the 500cc volume to
the device when prompted and then click
Next
.
Service manual
Functional verification and calibration overview 81
Summary of Contents for Spot Vital Signs 4400
Page 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Page 8: ...4 Symbols and definitions...
Page 34: ...30 Power up sequence...
Page 68: ...64 Field replaceable units...
Page 127: ...Interconnect diagram Service manual Appendix 123...
Page 128: ...124 Appendix...
Page 132: ......