Message
Possible cause
Suggested action
NIBP not functional. 05FF08
Sensor power up failure
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 05FF09
WACP rendezvous failure
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 05FF0A
Application firmware retrieval error
during POST
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 05FF0B
Upgrade .pim file is corrupt
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 05FF0C
Configured upgrade firmware
directory cannot be accessed
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 05FF0E
The NIBP sensor unexpectedly reset Clear the error and try again
NIBP not functional. 05FF0F
The NIBP sensor firmware failed to
upgrade
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
SpO2 messages
General SpO2 messages
Message
Possible cause
Suggested action
SpO2 not functional.
044800
SpO2 not functional
Internal hardware malfunction in SpO2
module. If the problem persists, have the
device inspected by a qualified service
technician.
SpO2 rebooting. 044900
The SpO2 module is not responding. Informational error. The host software is
attempting to clear an error by rebooting
the SpO2 module. No action required.
SpO2 rebooting. 044901
The SpO2 module stopped sending
data
Informational error. The host software is
attempting to clear an error by rebooting
the SpO2 module. No action required.
SpO2 rebooting. 044902
SpO2 received a packet with bad CRC
from module
Informational error. The host has received
a packet with bad CRC from the SpO2
module. The packet in question is
ignored. No action required.
SpO2 rebooting. 044903
The SpO2 power on self test failed
Internal hardware malfunction in SpO2
module. If the problem persists, have the
device inspected by a qualified service
technician.
110 Appendix
Summary of Contents for Spot Vital Signs 4400
Page 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Page 8: ...4 Symbols and definitions...
Page 34: ...30 Power up sequence...
Page 68: ...64 Field replaceable units...
Page 127: ...Interconnect diagram Service manual Appendix 123...
Page 128: ...124 Appendix...
Page 132: ......