NIBP specifications
Pulse rate range (using blood pressure determination)
Adult: 30 to 200 bpm (StepBP, SureBP)
Pediatric: 30 to 200 bpm (StepBP, SureBP)
Pulse rate accuracy (using blood pressure
determination)
±5.0% (±3 bpm)
Overpressure cutoff
Adult: 300 mmHg ±15 mmHg
Pediatric: 300 mmHg ±15 mmHg
SureTemp Plus temperature module specifications
SureTemp Plus temperature module specifications
Temperature range (all
measurement sites)
80°F to 110°F (26.7°C to 43.3°C)
Calibration accuracy
±0.2°F (±0.1°C) (Direct mode)
Laboratory accuracy
Temperature measurement range
Ambient
64.4°F (18°C) to 82.4°F
(28°C)
Ambient
50°F (10°C) to 64.4°F (18°C)
or 82.4°F (28°C) to 113°F
(40°C)
Less than 95.9°F (35.5°C)
±0.36°F (±0.2°C)
±0.36°F (±0.2°C)
95.9°F (35.5°C) to less than 96.4°F (35.8°C)
±0.18°F (±0.1°C)
±0.36°F (±0.2°C)
96.4°F (35.8°C) to less than 98.0°F (37.0°C)
±0.18°F (±0.1°C)
±0.3°F (±0.2°C)
98.0°F (37.0°C) to 102.0°F (39.0°C)
±0.18°F (±0.1°C)
±0.2°F (±0.1°C)
Greater than 102.0°F (39.0°C) to 106.0°F (41.0°C)
±0.18°F (±0.1°C)
±0.3°F (±0.2°C)
Greater than 106.0°F (41.0°C) to 107.6°F (42.0°C)
±0.18°F (±0.1°C)
±0.36°F (±0.2°C)
Greater than 107.6°F (42.0°C)
±0.36°F (±0.2°C)
±0.36°F (±0.2°C)
Thermometry clinical accuracy validation
For a copy of our Clinical Validation Study, please contact Hillrom Customer Care.
Service manual
Appendix 89
Summary of Contents for Spot Vital Signs 4400
Page 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Page 8: ...4 Symbols and definitions...
Page 34: ...30 Power up sequence...
Page 68: ...64 Field replaceable units...
Page 127: ...Interconnect diagram Service manual Appendix 123...
Page 128: ...124 Appendix...
Page 132: ......