Algorithm default
Select the default algorithm used to determine
NIBP measurements.
Cuff inflation target
If the default algorithm is set to Step, enter a default
cuff inflation pressure for each type of patient.
3. Optionally, configure other advanced settings.
4. When you are finished configuring advanced settings, touch
Exit
to return to the
Home
tab.
Specify advanced temperature settings
1. In Advanced settings, touch the
Parameters
>
Temperature
tabs.
2. Specify settings.
Setting
Action/Description
Default temperature site
Select a location for the default temperature site.
Unit of measure
Select primary units of measure for the temperature
display on the Home tab.
Display temperature conversion
Enable to display primary units of measure and
secondary units of measure for the temperature
display on the Home tab.
3. Optionally, configure other advanced settings.
4. When you are finished configuring advanced settings, touch
Exit
to return to the
Home
tab.
Specify additional parameters
1. In Advanced settings, touch the
Parameters
>
Additional parameters
tabs.
2. Select the additional parameters you wish to display on the
Home
tab.
NOTE
BMI will be automatically calculated based on the patient's
height and weight.
3. If you selected the options
Display height
and
Display weight
, then select the preferred unit
of measurement for each parameter.
4. Optionally, configure other advanced settings.
5. When you are finished configuring advanced settings, touch
Exit
to return to the
Home
tab.
Specify advanced program interval settings
1. In Advanced settings, touch the
Parameters
>
Averaging
tabs.
2. Specify settings.
Setting
Action/Description
Readings to average
The averaging program supports up to 6 readings.
Select which readings in the series to use for the
averaged measurement (and, as a result, the
number of readings to average).
Delay to start
Specify the amount of time to wait after the
clinician touches
Start Average
and before the
first reading starts.
Time between
Specify the amount of time to wait between each
reading.
Service manual
Advanced settings 25
Summary of Contents for Spot Vital Signs 4400
Page 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Page 8: ...4 Symbols and definitions...
Page 34: ...30 Power up sequence...
Page 68: ...64 Field replaceable units...
Page 127: ...Interconnect diagram Service manual Appendix 123...
Page 128: ...124 Appendix...
Page 132: ......