Message
Possible cause
Suggested action
Temperature not functional. 030814 The temperature module
TEMP CONFIG ACQUIRE
FAILURE
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Temperature not functional. 030815 The temperature module
TEMP CONFIG RELEASE
FAILURE
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Temperature not functional. 030816 The temperature module
TEMP CONFIG INVALID
PTR FAILURE
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Temperature not functional. 030817 EEPROM not initialized
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Unable to detect new temperature.
Retry measurement. 030818
The temperature module
heater indicates on when
turned off.
Probe malfunction. Replace probe. If the
problem persists, have the device inspected
by a qualified service technician.
Unable to detect new temperature.
Retry measurement. 030819
The temperature module
heater indicates off when
turned on.
Probe malfunction. Replace probe. If the
problem persists, have the device inspected
by a qualified service technician.
Temperature not functional. 03081A The temperature module
HTR_Q is on and HTRC is
off but still has voltage.
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Temperature not functional. 03081B The temperature module
HTR_Q is tri-stated with
HTRC enabled and has
heater power.
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Temperature not functional. 03081C The temperature module
turned Q&C on and the
heater voltage is not high
enough.
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Temperature not functional.
03081D
The temperature module
heater hardware failsafe
should have turned off
but did not.
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Replace temperature probe. 03081E The temperature module
probe is above 112°F or
43.3°C.
Probe malfunction. Replace probe. If the
problem persists, have the device inspected
by a qualified service technician.
Replace temperature probe. 03081F The temperature module
has excessive heater
energy
Probe malfunction. Replace probe. If the
problem persists, have the device inspected
by a qualified service technician.
Temperature not functional. 030820 The temperature module
host interface error
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Ambient temperature out of range.
Clear to retry. 030821
The temperature module
over ambient 45 °C temp
Verify conditions are less than 104 °F or 40 °C.
If conditions are valid and the problem
persists, replace the probe. If the problem
Service manual
Appendix 115
Summary of Contents for Spot Vital Signs 4400
Page 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Page 8: ...4 Symbols and definitions...
Page 34: ...30 Power up sequence...
Page 68: ...64 Field replaceable units...
Page 127: ...Interconnect diagram Service manual Appendix 123...
Page 128: ...124 Appendix...
Page 132: ......