Remove the LCD frame and display assembly
•
Separate the top housing assembly from the bottom housing assembly as described.
1. Remove the Pro-Gaff tape that holds the power switch flex cable to the LCD frame
and the top housing assembly.
2. Remove the six shoulder screws on the LCD frame using a #2 Phillips bit.
3. Remove the LCD frame.
4. Remove the LCD display and gasket from the top housing assembly.
Note
: Be sure to support the LCD display if you are removing it with the frame.
5. Remove the LCD display gasket.
Set aside the LCD gasket for reassembly.
6. Unlock the LCD flex connector from the LCD display assembly.
84 Disassembly, repair, and reassembly
CP 150 12-lead resting electrocardiograph
Summary of Contents for CP 150
Page 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Page 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Page 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Page 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Page 27: ...Home screen Home screen The Home screen includes the following areas 23...
Page 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Page 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Page 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Page 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Page 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Page 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Page 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Page 133: ......