3. Remove the two M3x5 screws from the ECG board using a #1 Phillips bit.
4. Push back slightly on the two latches to remove the ECG board from the bottom
assembly.
Reassembly note
•
Use a #1 Phillips bit and a Torque driver calibrated to 5.0 in-lb ± 0.5 in-lb to secure
the ECG board screws.
Remove the ECG ferrite
•
Separate the top housing assembly from the bottom housing assembly as described.
Gently open the two side-catches to remove the ECG ferrite and the ECG board
harness from the printer top assembly.
Note
Set the harness cable aside for reuse during reassembly.
88 Disassembly, repair, and reassembly
CP 150 12-lead resting electrocardiograph
Summary of Contents for CP 150
Page 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Page 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Page 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Page 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Page 27: ...Home screen Home screen The Home screen includes the following areas 23...
Page 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Page 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Page 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Page 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Page 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Page 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Page 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Page 133: ......