Reassembly notes
•
Use a #2 Phillips bit and a Torque driver calibrated to 5.0 in-lb ± 0.5 in-lb to secure
the LCD frame shoulder screws.
Ensure that the LCD flex cable is not under the LCD frame.
CAUTION
The LCD flex cable is extremely fragile and easily damaged. Do
not cause creases that may break the connections.
Insert the LCD display into the gasket with the exposed printed circuit board on the
bottom. Verify that the LCD flex cable feeds through the gasket's notched clearance
feature.
Place the LCD frame over the display. Verify that the LCD frame does not cover the LCD
flex cable.
Service manual
Disassembly, repair, and reassembly 85
Summary of Contents for CP 150
Page 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Page 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Page 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Page 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Page 27: ...Home screen Home screen The Home screen includes the following areas 23...
Page 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Page 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Page 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Page 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Page 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Page 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Page 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Page 133: ......