4. Disconnect the WiFi radio USB cable connector from the WiFi radio holder.
Reassembly notes
•
Use a #2 Phillips bit and a Torque driver calibrated to 5.0 in-lb ± 0.5 in-lb to secure
the two M4 pan head screws to the WiFi radio holder.
Remove the ECG board
•
Separate the top housing assembly from the bottom housing assembly as described.
1. Lift up the ECG cover to access the ECG board.
2. Disconnect the J401 connector from the ECG board.
Service manual
Disassembly, repair, and reassembly 87