4. Touch the
DICOM
tab to specify the DICOM settings.
Setting
Action/Description
Enable worklist downloads and ECG DICOM uploads
Select this option to enable DICOM functionality.
Local AE Title
Touch
and enter the AE Title for the device
(Example: CP150). Enter a maximum of 16 characters.
5. Touch
(Next) to view more DICOM Worklist Server settings. The DICOM
Worklist Server configuration settings appear.
Setting
Action/Description
Server AE Title
Touch
and enter the AE Title for the server.
Enter a maximum of 16 characters.
IP address
Touch
in the server IP address fields and enter the
IP address.
Port
Select the port. Touch
in the Port entry field and
enter the port number. (The port number is set by the
network administrator.)
Location filter
Use the drop-down menu to turn the filter off. Or filter by
Local AE Title or Device ID / Practice ID.
Test Connection
Touch
Test Connection
to test the connection to
the DICOM Worklist Server.
6. Touch
(Next) to view the DICOM Image Server settings. The DICOM Image
Server configuration settings appear.
7. Connect to the DICOM Image Server.
Setting
Action/Description
Server AE Title
Touch
and enter the AE Title for the DICOM
Image Server. Enter a maximum of 16 characters.
IP address
Touch
in the server IP address fields and enter the
IP address.
Port
Select the port. Touch
in the Port entry field and
enter the port number. (The port number is set by the
network administrator.)
ECG waveform storage
Select 12-Lead or General waveform storage format.
Note
When the selected ECG waveform is set to the 12-
Lead format, storage reverts to the General waveform if
the number of samples is larger than permitted for the
12-Lead format. The 12-Lead storage reverts to the
General waveform only if the Auto Report is set to one of
the 3x4 formats with the 5-second option and the ECG
test is for a pediatric patient.
Coding scheme
Select SCPECG or MDC.
Note
See the CP150 Conformance Statement for Coding
scheme definitions.
36 Settings
CP 150 12-lead resting electrocardiograph
Summary of Contents for CP 150
Page 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Page 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Page 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Page 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Page 27: ...Home screen Home screen The Home screen includes the following areas 23...
Page 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Page 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Page 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Page 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Page 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Page 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Page 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Page 133: ......