3. Remove the four M3x5 screws from the SMCE board using a #1 Phillips bit.
Reassembly note
•
Use a #1 Phillips bit and a Torque driver calibrated to 5.0 in-lb ± 0.5 in-lb to secure
the SMCE board M3x5 screws.
Remove the interface board
Note
This section only applies to 4th Edition electrocardiograms.
1. Lift the latch of the connectors on the SMCE board and the interface board and
disconnect the flex cables.
2. Remove the three screws on the interface board.