Printer pattern test and check point
3. Touch
Yes
if the test page printed correctly, or touch
No
if the test page did not print
correctly.
4. Touch (Next) to perform the next functional test.
Printer speed test
1. Touch each of the settings
10 mm/s
,
25 mm/s
, and
50 mm/s
to print a test page.
2. Use the speed test print page to verify the correct operation of the device printer by
comparing the pattern printed on the test page to point F in the
Printer speed test
and check point
graphic. The line should be a straight line that measures 100 mm
(+/-5 mm).
Printer speed test and check point
3. Touch
Yes
if the test page printed correctly, or touch
No
if the test page did not print
correctly.
4. Touch (Next) to perform the next functional test.
56 Service
CP 150 12-lead resting electrocardiograph
Summary of Contents for CP 150
Page 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Page 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Page 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Page 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Page 27: ...Home screen Home screen The Home screen includes the following areas 23...
Page 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Page 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Page 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Page 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Page 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Page 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Page 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Page 133: ......