USB host test
Note
Welch Allyn has not validated specific USB flash drives. Use only 64Gb or
less USB flash drives.
1. Insert a USB flash drive into each of the four ports on the back of the device and
follow the instructions on the screen to test each of the ports.
2. Touch
USB host test
to test each port from left to right.
3. Touch (Next) to perform the next functional test.
USB client test
1. Connect a USB type mini-B connector to the device client USB port and a type A
connector to a free USB port on the PC. Verify that the PC is able to recognize
CP150 device.
2. Touch
Yes
to confirm that your PC recognized the CP 150 device, or touch
No
if your
PC did not recognized the CP 150 device.
3. Touch (Next) to perform the next functional test.
52 Service
CP 150 12-lead resting electrocardiograph
Summary of Contents for CP 150
Page 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Page 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Page 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Page 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Page 27: ...Home screen Home screen The Home screen includes the following areas 23...
Page 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Page 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Page 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Page 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Page 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Page 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Page 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Page 133: ......