Service
The Service tab presents numerous settings and controls typically accessed by
authorized service or biomedical engineering personnel to configure, maintain, test, and
update the device. For example, the Service tab enables authorized users to save device
configurations to a USB flash drive and then load saved configurations to other devices.
Systems and devices configured with the PartnerConnect™ service feature also have
access to remote diagnostics, troubleshooting, and software upgrades.
View or change the Service settings
To view or change the settings
1. From the
Home
screen, touch the
Settings
tab.
2. Touch the
Advanced
tab.
3. Enter
6345
as the access code and touch
OK
. The General tab appears at the
bottom of the screen and the Regional tab appears at the top of the screen.
4. Touch the
Service
tab.
Restore factory defaults
Note
Your model might not contain all of these features.
•
All settings
•
Printer page count
•
Calibration gain
•
Radio settings
Device configuration
•
Save to USB
•
Configure from USB
•
Print all settings
Update software
•
Update
39
Summary of Contents for CP 150
Page 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Page 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Page 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Page 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Page 27: ...Home screen Home screen The Home screen includes the following areas 23...
Page 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Page 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Page 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Page 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Page 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Page 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Page 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Page 133: ......