Replacing the APPI lamp drive assembly O-rings
4-11
7. Use the 2.5-mm Allen wrench to unscrew the four captive lamp
mounting flange securing screws.
8. Remove the lamp mounting flange from the APPI source enclosure side
plate.
9. Use the O-ring removal kit to carefully remove the lamp mounting
flange O-ring from the APPI source enclosure side plate.
“Removing O-rings and seals” on page 4-3
.
Captive securing screw (4-off)
Lamp mounting flange
Lamp mounting flange O-ring
APPI source enclosure side plate