Considerations specific to the Quattro Premier XE Combined
APPI and APCI Source
Solvent leakage hazard
The source exhaust system is designed to be robust and leak-tight. Waters
recommends you perform a hazard analysis, assuming a maximum leak into
the laboratory atmosphere of 10% LC eluate.
Warning:
• To confirm the integrity of the source exhaust system, renew
the source O-rings at intervals not exceeding one year.
• To avoid chemical degradation of the source O-rings, which
can withstand exposure only to certain solvents (see
“Materials of Construction and Compliant Solvents” on
), and determine whether any solvents you use that
are not listed are chemically compatible with the composition
of the O-rings.